Paper Abstract and Keywords |
Presentation |
2020-11-20 11:15
A small imaging system for dual-area imaging by CMOS imaging sensors Ronnakorn Siwadamrongpong, Kenji Sugie, Makito Haruta, Hironari Takehara, Hiroyuki Tashiro, Kiyotaka Sasagawa, Jun Ohta (NAIST) |
Abstract |
(in Japanese) |
(See Japanese page) |
(in English) |
In this study, we developed a small circuit system to operate implantable imaging devices based on a complementary metal-oxide-semiconductor (CMOS) image sensor. This device could achieve low invasiveness to the animal and make multi-area implantation possible. A microcontroller embedded with the USB high-speed interface is used in the design of the system with a size of 30 × 40 mm2. The imaging experiment with the 124 × 269 Pixel CMOS image sensors showed an average frame rate of 80 fps in a dual-sensor setup. |
Keyword |
(in Japanese) |
(See Japanese page) |
(in English) |
CMOS image sensor / Microcontroller / USB high speed / Dual-area imaging / / / / |
Reference Info. |
ITE Tech. Rep., vol. 44, no. 29, IST2020-46, pp. 13-15, Nov. 2020. |